NOE Experiments
When a proton is saturated or
inverted, spatially-close protons may experience an intensity enhancement, which
is termed the Nuclear Overhauser Effect (NOE). The NOE is unique among NMR
methods because it does not depend upon through-bond J couplings but depends
only on the spatial proximity between protons. In other words, the strength of the NOE gives information on
how close two protons are. For
small molecules, an NOE may be observed between protons that are up to 4Å
apart, while the upper limit for large molecules is about 5Å.
Molecular Weight and
Maximum NOE
The maximum possible NOE
depends on the molecular correlation time (or the inverse of the rate of
molecular tumbling), which is in large part determined by the molecular weight
and solvent viscosity. Larger
molecular weights and higher viscosities lead to larger correlation times. The NOE is positive for small molecules
(MW< 600), goes through zero for medium-sized molecules (MW range 700
– 1200), and becomes negative for large molecules (MW>1200). (These MW ranges are approximate only.) For medium sized molecules, the NOE
may be theoretically zero. See
the figure below that is adapted from Newhaus and Williamson text. The ROESY
experiment (rotating frame NOE) is preferred for medium-sized molecules since
the ROE is always positive.

Time Dependence of NOE -
Mixing Times
In transient experiments,
such as NOESY and ROESY, the NOE dynamically builds up and then decays due to
relaxation during the mixing time, as shown below in the plot of NOE versus
mixing time. The NOE, thus, goes through a maximum as function of mixing
time. The location of the maximum
NOE and rate of build-up depend on the correlation time, or its proxy, the
molecular weight, and the distance between protons for a particular NOE. In general, large molecules build-up
NOE quickly while small molecules build-up NOE more slowly. That is, for large
molecules the point of maximum NOE is shifted to shorter mixing times. A shorter distance between protons will
also lead to faster build-up of NOE and a shift of the maximum to shorter
mixing times.
There is only one mixing time
specified per NOE experiment, and it is the most important parameter for NOE
experiments. For small molecules, a mixing time that maximizes the NOE is
desirable, unless you intend to calculate an actual distance (see analysis
section). Generally, one is
interested in a range of distances so the choice depends on molecular weight
rather than a particular distance.
For large molecules, the mixing time must be kept small so that the build-up
obeys the linear approximation and spin diffusion is avoided (see analysis
section). The following are guidelines:

1) small molecules 0.5 -1 sec. Start with 0.5 sec.
2) medium size molecules 0.1 -0.5 sec. Start with 0.25 sec.
3) large molecules 0.05 - 0.2
sec. Start with 0.1 sec.
1D versus 2D Methods
The choice between 2D (ROESY
or NOESY) versus 1D (selective NOESY) depends on the amount of material
available and the amount of information needed. A single 2D experiment gives all NOE information
simultaneously whereas 1D experiments provide NOEs one at a
time. In general, I
recommend the 2D methods. The minimum amount of time required
(which does not depend on sample concentration but on the time necessary for experimental
cycling) for 2D and 1D differ.
The minimum time for a 2D NOESY spectrum is longer. The standard 2D NOESY often requires a
minimum of 1.5 hours but the 2D NOESY with zero-quantum suppression, which uses
gradients, has a minimum time of only 25 minutes. The minimum time for a single
1D selective NOESY spectrum is about 2 minutes. Many 1D experiments, however, are usually required. If you
have very little material, then signal averaging will be required anyway and
the 2D version should be used.
Spectral crowding will affect
the choice of experiment. If
critical peaks to be irradiated are very close (<30 Hz) to other peaks, then
the selectivity of the 1D version will not be sufficient and the 2D version
will be needed.
Artifacts and Their Suppression
Zero-quantum peaks are a
common artifact in all NOESY spectra.
They occur between peaks that are J-coupled, such as ortho-protons on a
ring, as can be identified by their up-down DQF-COSY type of pattern.
There is a 2D NOESY sequence
that is designed to remove these zero-quantum peaks.
In ROESY spectra, a common
occurrence is TOCSY transfer between protons that are J-coupled or symmetric
with respect to the center of the spectrum. This latter artifact can be removed by proper positioning of
o1p, the center of the spectrum. Finally, the cross-peak intensities have an
offset dependence. See analysis section for more detail.
If protons are undergoing
chemical exchange, corresponding cross peaks occur in all NOE and ROE
experiments. In fact, chemical
exchange can be studied with these same NOE methods and are then termed EXESY
experiments.
In 1D selective NOESY
experiments, there are several types of possible artifacts: zero-quantum
(up-down) peaks as well as the unsuppressed residual from very intense
singlets. Moreover, the
experiment uses selective pulses and their proper calibration is required for
optimal results and suppression of artifacts.
Choice of Experiment
– a Prescription
Small molecules (MW < 600)
The usual choice is 2D NOESY with zero-quantum suppression. Exceptions would be if you have a very concentrated sample and you are only interested in one or two NOEs and the peaks to be irradiated are well-separated; then choose the 1D selective NOESY. ROESY has only disadvantages for small molecules.
Medium sized molecules (700
< MW < 1200)
ROESY
is preferred.
Large Molecules (MW >
1200)
The
choice here is more complicated.
The usual choice is 2D NOESY but 2D ROESY has advantages. ROESY suffers less from spin diffusion
and the resulting interpretation errors.
However, ROESY is less sensitive for large molecules and has other
disadvantages such as TOCSY artifacts. See analysis section.
Sample Considerations:
Dissolved oxygen or other
paramagnetic species such as Cu2+ can reduce or completely quench
the NOE. For small molecules, it is extremely
important to remove dissolved oxygen.
For large molecules,
the removal of oxygen is not critical. Removal of oxygen must be done by the freeze-pump-thaw
method. Simply bubbling argon
through the sample is not sufficient. The following describes the freeze-pump-thaw
procedure:
1) freeze the sample in
liquid nitrogen or CO2/acetone.
2) evacuate the space above
the solution.
3) turn off vacuum but keep
sample isolated and allow to thaw.
As it thaws, bubbling should be noticed.
4) repeat several times (3-4
times).
5) backfill with N2.
When finished, the sample
should, of course, be sealed in some manner. Tubes with attached stopcocks are available.
Sample size and tube options
When sample quantity is very
limited, it is advantageous to limit the amount of solvent in which it is
dissolved. If a normal 5mm tube is
used, however, this cannot be less than about 500mL without causing serious lineshape problems (shimming
problems) and the attendant loss of signal-to-noise. There are special tubes made by Shigemi, however, that
can be used to restrict the active volume and, hence, reduce the amount of
solvent without causing lineshape problems. Shigemi tubes are available from
Aldrich.
ANALYSIS – peak
identification
Relative Phase of Cross Peaks
The
phase of ROE, NOE and chemical
exchange cross peaks can be different and are summarized in the figure below.
In this figure, it is assumed that protons A and B have an NOE or ROE while
protons C and D are undergoing chemical exchange. Note that the phase behavior differs for large and small
molecules. For small molecules, the diagonal peaks and NOE cross peaks have
opposite phase. If the diagonal is negative, then NOE cross peaks will be
positive. For large molecules, the
diagonal and the NOE cross peaks have the same phase. The phase of the cross peaks, then, indicates whether
the molecule is in the large or small molecule region, which has important
implications for quantitation, as discussed below. Note that this phase behavior is due to the
positive/negative nature of NOE described at the beginning of this handout.
Cross
peaks due to chemical exchange, if it is occurring, have the same phase as the
diagonal for both small and large molecules in both ROESY and NOESY.
In
ROESY, the diagonal peaks and ROE cross peaks have opposite phase for all
molecules since the ROE is always positive. TOCSY cross peaks are the major artifact in ROESY
spectra. TOCSY peaks have the same
phase as the diagonal, and are thus similar to exchange peaks. TOCSY occurs between spins that are J
coupled and that are relatively close in chemical shift. It also occurs for peaks that are
symmetric about o1p. A possible
complication is the relay of ROE through TOCSY resulting in false ROESY cross
peaks. For example, geminal methylene peaks often show TOCSY cross peaks. Assume there is a third proton that
should have an ROE to only one of the geminal protons but not its partner. TOCSY can transfer the ROE to its
partner and it may appear as if the third proton has an ROE to both geminal
protons.

Spin Diffusion
When analyzing NOESY spectra,
one must understand the consequences of spin diffusion. Spin diffusion occurs
primarily for large molecules and for long mixing times outside the “linear
approximation”. In NOESY
spectra, spin diffusion can lead to misleading cross peaks and incorrect
distances. In this section, I
describe the presence of extra, misleading cross peaks and in the quantitation
section, I discuss incorrect distances. Assume there are four protons A, B, and
E and F and that A and B, B and E, and E and F are close. That is, you expect NOEs between those
three pairs, as shown in the figure below. These expected cross peaks between protons that are close
are termed direct contributions.
When spin diffusion is present, indirect contributions will also be
present and a cross peak between A and C will likely be present. In spin diffusion, the magnetization
follows a path from A to B and then from B to C but appears to be directly from
A to C. In NOESY spectra of large molecules, the phase of these indirect peaks
is the same as for direct contributions and the resulting cross peaks are
impossible to distinguish at a single mixing time. The appearance of the A to C
cross peak could lead you to erroneously conclude that protons A and C are
close.
ROESY spectra suffer much
less from spin diffusion; the phase of indirect contributions may be different
from direct contributions and allows their easy identification. The phase of indirect contributions
alternates with number of steps of transfer. That is, the phase of 2-step indirect contributions is
opposite to direct contributions, while that of 3-step indirect contributions
is the same as direct contributions. 3-step contributions are rare,
however. (see Bax, J. Magn. Res. 70, 327-331 (1986))

For organic molecules, it is
generally sufficient to classify NOE peak intensities as strong, medium, and
weak and make qualitative deductions about relative distances. If an actual distance is needed,
one may use the well-known approach in which the NOE is inversely proportional
to the distance to the 6th
power, i.e.,
rij= rref (aref/aij)1/6
where aij is the NOE cross-peak volume and rij is the interproton distance of the the two protons i and j. Given a known distance between two protons (rref) and its NOE volume (aref), a distance can calculated from another NOE volume.
For this relation to be
valid, a strict experimental protocol must followed. First, the mixing time must be relatively short so
that the linear approximation is valid and spin diffusion is avoided. For small
molecules, the mixing time must be less than several hundred milliseconds. For large molecules, there may be no
practical value for a mixing time that completely avoids spin diffusion but, in
general, the mixing time must be less than 100 msec. Whether spin diffusion leads to an apparent increase or
decrease in distance depends on the details of the molecular geometry. Linear
geometries lead to shorter apparent distances while non-linear geometries may
lead to longer distances (See
Neuhaus and Williamson p117-122) To help ensure that the mixing time is within
the linear region, a build-up curve is performed. A build-up curve is a series
of NOE spectra taken at different mixing times. If one is within the linear region, then the NOE will
linearly increase with mixing time.
A second requirement for quantitative work is that the relaxation
delay must be long enough to allow reasonable recovery of the magnetization
between scans. The normal time of
2 sec for D1 is not sufficient and one must increase this. The proper D1
should be at least 3 times T1. A T1 determination may be
necessary.
![]()
the effective distance for
NOE is less than the average distance, reff < rave.
and the effective distance is weighted towards that of the “closest approach”. For example, assume there are two
conformations A and B and that the A conformation is 10% populated. If the protons are 0.2 nm apart in
conformation A and 0.6 nm apart in conformation B, then the effective distance
is 0.293 nm which is much closer than the 0.6 nm separation that is present in
the dominant conformer. This is
summarized in the table below. Thus, in such cases the NOE will reflect the
conformation where the protons are closer.

ROESY – Quantitative Distance Determination
In addition to the above
considerations for NOESY, the ROESY has additional complications. The cross peak intensities have
an offset dependence relative to the transmitter center, o1p. Cross-peaks are less intense the
further they are from the center, regardless of spatial distance. For example,
assuming o1p is 5 ppm, then a cross-peak between protons at 1 and 2 ppm will
have lower intensity than between protons at 4 and 5 ppm, even if they have the
same interproton distance.
This dependence is well characterized and can be corrected in the
following way (see Ammalahti, et al.. J. Magn. Res. A, 122, 230-232
(1996)). Distances are calculated
from corrected intensities:
rij= rref (arefcref/aijcij)1/6
where
cij= 1/(sin2qisin2qj)
and
tanqI = gB1/(wI -w0)
where (wI -w0) is the difference between the chemical
shift of the peak (in Hz) and o1p (in Hz) and gB1 is the spin lock power (which is about
2500 Hz in our case). Volume
corrections of up to a factor of 4, in far off-resonance cases, may be required.
An additional complication with quantitation of ROESY spectra is that TOCSY transfer may occur and cancel or partially cancel ROESY cross peaks. This obviously has deleterious effects on distance determination. This is a particular problem for the reference ROE for which a J-coupled methylene pair is often chosen. See John Decatur for advice if detailed distance information is needed.